TY - JOUR
T1 - Computerized method for nonrigid MR-to-PET breast-image registration
AU - Unlu, M. Z.
AU - Krol, A.
AU - Magri, A.
AU - Mandel, J. A.
AU - Lee, W.
AU - Baum, K. G.
AU - Lipson, E. D.
AU - Coman, I. L.
AU - Feiglin, D. H.
N1 - Funding Information:
This work was partially supported by grants from the Carol M. Baldwin Breast Cancer Research Foundation of Central New York, and from the Department of Radiology, SUNY Upstate Medical University. M.Z. Unlu was supported by a grant from Turkish Ministry of National Education and by Syracuse University.
PY - 2010/1
Y1 - 2010/1
N2 - We have developed and tested a new simple computerized finite element method (FEM) approach to MR-to-PET nonrigid breast-image registration. The method requires five-nine fiducial skin markers (FSMs) visible in MRI and PET that need to be located in the same spots on the breast and two on the flanks during both scans. Patients need to be similarly positioned prone during MRI and PET scans. This is accomplished by means of a low gamma-ray attenuation breast coil replica used as the breast support during the PET scan. We demonstrate that, under such conditions, the observed FSM displacement vectors between MR and PET images, distributed piecewise linearly over the breast volume, produce a deformed FEM mesh that reasonably approximates nonrigid deformation of the breast tissue between the MRI and PET scans. This method, which does not require a biomechanical breast tissue model, is robust and fast. Contrary to other approaches utilizing voxel intensity-based similarity measures or surface matching, our method works for matching MR with pure molecular images (i.e. PET or SPECT only). Our method does not require a good initialization and would not be trapped by local minima during registration process. All processing including FSMs detection and matching, and mesh generation can be fully automated. We tested our method on MR and PET breast images acquired for 15 subjects. The procedure yielded good quality images with an average target registration error below 4 mm (i.e. well below PET spatial resolution of 6-7 mm). Based on the results obtained for 15 subjects studied to date, we conclude that this is a very fast and a well-performing method for MR-to-PET breast-image nonrigid registration. Therefore, it is a promising approach in clinical practice. This method can be easily applied to nonrigid registration of MRI or CT of any type of soft-tissue images to their molecular counterparts such as obtained using PET and SPECT.
AB - We have developed and tested a new simple computerized finite element method (FEM) approach to MR-to-PET nonrigid breast-image registration. The method requires five-nine fiducial skin markers (FSMs) visible in MRI and PET that need to be located in the same spots on the breast and two on the flanks during both scans. Patients need to be similarly positioned prone during MRI and PET scans. This is accomplished by means of a low gamma-ray attenuation breast coil replica used as the breast support during the PET scan. We demonstrate that, under such conditions, the observed FSM displacement vectors between MR and PET images, distributed piecewise linearly over the breast volume, produce a deformed FEM mesh that reasonably approximates nonrigid deformation of the breast tissue between the MRI and PET scans. This method, which does not require a biomechanical breast tissue model, is robust and fast. Contrary to other approaches utilizing voxel intensity-based similarity measures or surface matching, our method works for matching MR with pure molecular images (i.e. PET or SPECT only). Our method does not require a good initialization and would not be trapped by local minima during registration process. All processing including FSMs detection and matching, and mesh generation can be fully automated. We tested our method on MR and PET breast images acquired for 15 subjects. The procedure yielded good quality images with an average target registration error below 4 mm (i.e. well below PET spatial resolution of 6-7 mm). Based on the results obtained for 15 subjects studied to date, we conclude that this is a very fast and a well-performing method for MR-to-PET breast-image nonrigid registration. Therefore, it is a promising approach in clinical practice. This method can be easily applied to nonrigid registration of MRI or CT of any type of soft-tissue images to their molecular counterparts such as obtained using PET and SPECT.
KW - FEM-based soft tissue multimodality nonrigid image registration
KW - MR-to-PET nonrigid breast-image registration
UR - http://www.scopus.com/inward/record.url?scp=74149086948&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=74149086948&partnerID=8YFLogxK
U2 - 10.1016/j.compbiomed.2009.10.010
DO - 10.1016/j.compbiomed.2009.10.010
M3 - Article
C2 - 19942214
AN - SCOPUS:74149086948
SN - 0010-4825
VL - 40
SP - 37
EP - 53
JO - Computers in Biology and Medicine
JF - Computers in Biology and Medicine
IS - 1
ER -